Products DC Laboratory Power Supplies DP-M SERIES HIGH-PERFORMANCE POWER SUPPLIES DP-P SERIES MULTI-FUNCTIONAL LABORATORY POWER SUPPLIES DP-PH SERIES HIGH-VOLTAGE POWER SUPPLIES DP-S SERIES COMPACT LABORATORY POWER SUPPLIES DP-H SERIES HIGH PRECISION LABORATORY POWER SUPPLIES DP-HA SERIES AUTORANGING LABORATORY POWER SUPPLIES AC Laboratory Power Supplies DF-C SERIES AC INDUSTRIAL POWER SUPPLIES DF-CM SERIES HIGH-FREQUENCY AC POWER SUPPLIES DF-S SERIES AC LABORATORY POWER SUPPLIES

DP-HA Series AUTORANGING LABORATORY POWER SUPPLIES

Maximum values for the series (depending on model): 605 V / 111 A / 1.5 kW / ØPL = 1.79 €/W

DP-H Precision DC Power Supply – Front View
Voltage range: 0 ... 605V Setting accuracy U: ±0.01% of reading + 10 mV (35 V / 80 V / 85 V)
±0.03% of reading + 100 mV (155 V)
±0.03% of reading + 200 mV (305 V / 605 V)
Current range: 0 ... 111A Setting accuracy I: ±0.1% reading + 10mA
±0.1% reading + 60mA (111A)
Power range: 0 ... 1500W Ext. control analogue: No
Cabinet: Table-top unit Ext. feedback analogue: No
Regulation: Auto-ranging U / I RS232: Yes
Load regulation U: Depending on the model: ≤0.01% + 5 mV to ≤0.01% + 100 mV RS485: Yes
Load regulation I: Depends on the model: ≤0.1% + 5 mA to ≤0.1% + 10 mA RS422: No
Mains regulation U: Depends on the model: ≤0.01% + 2.5 mV to ≤0.01% + 50 mV Open protocol: Yes
Line regulation I: Depending on the model: ≤0.1% + 2.5 mA to ≤0.1% + 10 mA Modbus RTU: Yes
Residual ripple U: Depending on the model: ≤50 mVpp to ≤300 mVpp SCPI: Yes
Residual Ripple I: Depending on the model: ≤15 mArms to ≤150 mArms Reverse current protection: Yes
Recovery time (10% → 90% load): ≤5ms Polarity changer: No
Display accuracy U: ±0.01% of reading + 5 mV (35 V / 80 V / 85 V)
±0.02% of reading + 50 mV (155 V)
±0.02% of reading + 100 mV (305 V / 605 V)
Interlock: No
Display accuracy I: ±0.1% reading + 0.1%FS
±0.1% reading + 40mA (111A)
External Output On/Off: Yes
External program trigger: Yes Remote measurement: Up to 2V
Voltage range: 0 ... 605V
Current range: 0 ... 111A
Power range: 0 ... 1500W
Cabinet: Table-top unit
Regulation: Auto-ranging U / I
Load regulation U: Depending on the model: ≤0.01% + 5 mV to ≤0.01% + 100 mV
Load regulation I: Depends on the model: ≤0.1% + 5 mA to ≤0.1% + 10 mA
Mains regulation U: Depends on the model: ≤0.01% + 2.5 mV to ≤0.01% + 50 mV
Line regulation I: Depending on the model: ≤0.1% + 2.5 mA to ≤0.1% + 10 mA
Residual ripple U: Depending on the model: ≤50 mVpp to ≤300 mVpp
Residual Ripple I: Depending on the model: ≤15 mArms to ≤150 mArms
Recovery time (10% → 90% load): ≤5ms
Display accuracy U: ±0.01% of reading + 5 mV (35 V / 80 V / 85 V)
±0.02% of reading + 50 mV (155 V)
±0.02% of reading + 100 mV (305 V / 605 V)
Display accuracy I: ±0.1% reading + 0.1%FS
±0.1% reading + 40mA (111A)
Setting accuracy U: ±0.01% of reading + 10 mV (35 V / 80 V / 85 V)
±0.03% of reading + 100 mV (155 V)
±0.03% of reading + 200 mV (305 V / 605 V)
Setting accuracy I: ±0.1% reading + 10mA
±0.1% reading + 60mA (111A)
Ext. control analogue: No
Ext. feedback analogue: No
RS232: Yes
RS485: Yes
RS422: No
Open protocol: Yes
Modbus RTU: Yes
SCPI: Yes
Reverse current protection: Yes
Polarity changer: No
Interlock: No
External Output On/Off: Yes
External program trigger: Yes
Remote measurement: Up to 2V

* Optional

The DP-HA Series laboratory power supplies offer a flexible output range for varying test tasks thanks to auto-ranging. Within the rated power range, you can use both higher voltages at lower currents and lower voltages at higher currents. This allows you to cover different operating points with the same power supply.

You define recurring test sequences yourself: Voltage, current, and power waveforms are set over time, stored in the device, and executed automatically. The device is operated directly from the front panel; no external computer is required to run the sequence. A battery charging function is also available.

RS232 with SCPI and RS485 with Modbus RTU are provided as standard for integration into computer-controlled test benches. The remote measurement compensates for voltage drops in the supply line of up to 2 V.

Connectivity RS232 with SCPI and RS485 with Modbus RTU are standard. External trigger for programmed sequences. No analog setpoint or actual value interfaces.
Features Autoranging, CC/CV mode with automatic switching, user-defined test sequences, and battery charging function. Remote measurement to compensate for line losses up to 2 V.
Output configuration Potential-free, short-circuit-proof output. Maximum voltage and maximum current can only be combined within the rated power limit.
Security Overtemperature protection, as well as adjustable and activatable overvoltage and overcurrent protection with a status indicator. Lockable front-panel controls. The program input is a trigger and not a safety interlock.
Upgrade options Application-specific extension via AP9 and AP8.

Compact performance

DP-HA Auto-ranging DC Power Supply with Carrying Handle, Front View

Autoranging in a compact design

Illustration of a DP-HA laboratory power supply.

The compact design with a carrying handle makes it easy to move the unit between test stations. A temperature-dependent fan control system with five temperature sensors adjusts the cooling to the operating conditions.

Four devices in a rack mount

Removable rubber pads protect the front and back from physical impact. Without these pads, up to four units can be mounted in a 4U rack insert (sold separately). This creates a space-saving setup with four independently usable channels.

Programming and Monitoring

4.3-inch color display and front-panel controls
The color display shows setpoint and actual values as well as the time history of the output variables. A numeric keypad and a rotary encoder allow for direct value entry and menu navigation. The setting and display resolution is 1 mV / 1 mA or 10 mV / 10 mA, depending on the model. The accuracy is specified separately in the technical data.

Custom Time-Dependent Test Sequences
Create your voltage, current, and power waveforms to suit the test task and save the sequences in the device. A program can contain up to 100 steps. Specify the number of repetitions and start execution from the front panel or via the external trigger. Use a USB drive to transfer saved sequences to other devices.


Battery Charging Function: The programmable charging function allows you to specify an application-specific charging profile. Voltage and current can also be logged via an external controller. The charging parameters and protective measures must be compatible with the connected battery.

Autoranging: Example from the Series—
As an example from the series, the DP155-20-1500HA model provides a maximum of 155 V, 20 A, and 1,500 W. Within the rated power of 1,500 W, up to 75 V at 20 A or approximately 9.68 A at 155 V are possible. Autoranging expands the available output combinations but does not increase the rated power.

Dynamics and Specifications
The recovery time is ≤5 ms for a load change from 10% to 90%. Residual ripple and accuracy depend on the model. RMS and peak-to-peak values are not equivalent. The measurement value specified in the product data sheet is the relevant one in each case.

AP Series Expansion Modules

Front view of a DP-M high-performance DC power supply

Application-Specific Extensions

Illustration of a DP-M power supply with an AP8 module. The design of the expansion is tailored to the corresponding power supply and the application.

Integrated reverse current protection and additional protection modules
The output rectifier stage provides basic protection against reverse currents of the same polarity within the device’s permissible limits. This does not imply unlimited reverse current capability or suitability for absorbing arbitrary load energy. Reverse voltages, high reverse currents, and the stored energy of the load must be taken into account during design.

Model AP9: Application-Specific Reverse Current and Reverse Voltage Protection
The AP9 complements the integrated protection for demanding inductive or capacitive loads. The module is designed to order for the expected currents, voltages, and energy levels. Depending on the application, a combined protection circuit against high reverse currents and reverse voltages—particularly against reverse-polarity transients—or a design as an active braking chopper is possible. A braking chopper limits the energy fed back into the system by converting it into heat. The design, sizing, and permissible duty cycles are agreed upon on a case-by-case basis.

Model AP8: Additional output-side EMC filter with capacitance
AP8 combines a three-stage EMC filter with additional output capacitance. It serves to further reduce line-borne interference and stabilize the power supply to demanding loads. The design is tailored to the power supply unit, load, and dynamic requirements.

All AP modules are manufactured to order, with their enclosures and electrical designs customized to meet the agreed-upon requirements.

Connection options

Connections on the back of a DP-HA Autoranging DC Power Supply

Digital Control and External Program Triggers

Schematic illustration of the connections on a DP-HA laboratory power supply. The position and design may vary depending on the revision.

RS232 and RS485 (Standard)
Communication takes place via RS232 using SCPI and via RS485 using Modbus RTU. The documented protocols and included LabVIEW examples make it easier to develop your own control solutions.

Remote Measurement Input (Standard)
The remote measurement input measures the voltage directly at the load and allows for compensation of line losses up to 2 V. This is particularly helpful for high currents and longer connection cables.

External Trigger (Standard)
The trigger input is used to start or stop stored automatic sequences. The program trigger is not a safety interlock. An interlock loop is not provided as standard or as an option in this series. The DP-HA series does not include analog inputs or outputs for setting or monitoring voltage or current.

ModelPriceOutput voltageOutput currentPowerDisplay resolution UDisplay resolution IResidual ripple UResidual ripple ILoad regulation ULoad regulation ILine regulation ULine regulation I
DP155-20-1500HA2.210,00 €0 ... 155V0 ... 20A0 ... 1.5kW10mV1mA≤150mVpp≤30mArms≤0.01% + 40mV≤0.1% + 10mA≤0.01% + 30mV≤0.1% + 10mA
DP155-20-850HA1.620,00 €0 ... 155V0 ... 20A0 ... 850W10mV1mA≤150mVpp≤30mArms≤0.01% + 40mV≤0.1% + 10mA≤0.01% + 30mV≤0.1% + 10mA
DP305-10-1500HA2.210,00 €0 ... 305V0 ... 10A0 ... 1.5kW10mV1mA≤250mVpp≤40mArms≤0.01% + 100mV≤0.1% + 10mA≤0.01% + 50mV≤0.1% + 10mA
DP305-10-850HA1.670,00 €0 ... 305V0 ... 10A0 ... 850W10mV1mA≤250mVpp≤40mArms≤0.01% + 100mV≤0.1% + 10mA≤0.01% + 50mV≤0.1% + 10mA
DP35-111-1500HA2.180,00 €0 ... 35V0 ... 111A0 ... 1.5kW1mV10mA≤100mVpp≤150mArms≤0.01% + 5mV≤0.1% + 10mA≤0.01% + 5mV≤0.1% + 10mA
DP35-111-850HA1.590,00 €0 ... 35V0 ... 111A0 ... 850W1mV10mA≤100mVpp≤150mArms≤0.01% + 5mV≤0.1% + 10mA≤0.01% + 5mV≤0.1% + 10mA
DP605-5-1500HA2.300,00 €0 ... 605V0 ... 5A0 ... 1.5kW10mV1mA≤300mVpp≤30mArms≤0.01% + 100mV≤0.1% + 10mA≤0.01% + 50mV≤0.1% + 10mA
DP605-5-850HA1.730,00 €0 ... 605V0 ... 5A0 ... 850W10mV1mA≤300mVpp≤30mArms≤0.01% + 100mV≤0.1% + 10mA≤0.01% + 50mV≤0.1% + 10mA
DP80-20-400HA1.040,00 €0 ... 80V0 ... 20A0 ... 400W1mV1mA≤50mVpp≤15mArms≤0.01% + 5mV≤0.1% + 5mA≤0.01% + 2.5mV≤0.1% + 2.5mA
DP85-40-1500HA2.210,00 €0 ... 85V0 ... 40A0 ... 1.5kW1mV1mA≤100mVpp≤50mArms≤0.01% + 10mV≤0.1% + 10mA≤0.01% + 10mV≤0.1% + 10mA
DP85-40-850HA1.620,00 €0 ... 85V0 ... 40A0 ... 850W1mV1mA≤100mVpp≤50mArms≤0.01% + 10mV≤0.1% + 10mA≤0.01% + 10mV≤0.1% + 10mA